Do You Know How to Maintain Your Luminex® 100/200™ Instrument?

Luminex 200 Instrument

Does instrument maintenance sometimes fall off your calendar? The xPONENT® 3.1 software makes maintaining your Luminex® 100/200™ instrument quick and easy.

To keep your unit running smoothly year-round, follow the instructions provided in the instrument and software user manuals. However, this is a quick guide for maintaining your instrument.

If you are not currently running xPONENT software, contact your Business Manager or Technical Support to receive information on a software upgrade. You will be glad you did!

The Daily and Weekly Schedule:

NOTE: When 10% bleach is required, make sure you are using fresh bleach.

Daily:

  • Perform “System Initialization” protocol:
    • Laser Warm-up, Fluidics, and Performance Verification
    • Note: The tasks included in the System Initialization can be changed as follows:
      • Admin page/System Setup tab, adjust System Initialization settings, use pull-down menu
  • Perform “System Shutdown” protocol at the end of the day

Weekly:

  • Clean probe (sonicate >2 min, flush probe from small end)
  • Run Self-Test
    • On Maintenance page/Cmds & Routines tab
    • Use drop down menu; select None and choose “Self Test”
  • Run “Weekly Maintenance” protocol
    • On Maintenance page/Cmds & Routines tab
    • Use drop down menu
    • Set to “Laser warm-up, Fluidics, Calibration, and Performance Verification” (See above for changing the tasks included in the System Initialization)
  • Perform Calibration, change “System Initialization” protocol:
    • Set to “Laser warm-up, Fluidics, Calibration, and Performance Verification” (See above for changing the tasks included in the System Initialization)

Bi-Annual and Annual Schedule:

  • See “Schedule” tab under the “Maintenance” page.

As Needed Maintenance:

 
If you are not going to use the instrument for over a week:

  • Run the “Prepare for Storage #1” and “Prepare for Storage #2” routine. When you go to run the instrument again, run the “Revive from Storage” routine.
    • Routines are found on the Maintenance page/Cmds & Routines tab (use the drop down menu).
    • When you get past about a week, the water held in the probe from the last set of commands has likely evaporated and possibly left a salt residue.
    • If this happens just once or twice, you most likely won’t notice any issues. However, when the probe builds up the salt residue over time (many cycles of non-use) bead count issues start to arise.

Clogs/Fluidics failures:

  • Try cleaning the probe as stated under weekly maintenance. See “How Well Do You Know Your Sample Probe?” blog.
    • Note: Increased probe cleaning may be needed in the following events: calibration failure, slow event rate is detected, sample empty errors, or fluid dripping from the probe
  • You can also try a sodium hydroxide cleaning procedure.
  • Do an extra cleaning routine after every assay to help avoid the buildup in the probe that can happen when using serum samples:
    • This protocol can be created in xPONENT by clicking on “Maintenance,” “Commands and Routines,” and while “None” is selected in the dropdown, create the following routine and save it:
      • 1 x Sanitize with 10% bleach or 0.1N NaOH
      • 1 x –Alcohol Flush
      • 1 x Wash with diH2O

Probe height adjustment:

Refer to your xPONENT Software User Manual for step-by-step instructions.

This should give you a good start, but for more in-depth, step-by-step instructions, refer to your xPONENT Software User Manual. You can also find a number of maintenance and help videos on our website. Find them here!